Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8952405 | Brachytherapy | 2018 | 6 Pages |
Abstract
Both the RABBIT tool and process mapping and FMEA were shown to be valuable tools in developing a QM program for ultrasound-based HDR pBT treatments. A considerable number of the potential failure modes identified in both tools were related to human or procedural errors, highlighting the importance of checklists and protocols in delivering a safe and effective ultrasound-based HDR pBT treatment.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Oncology
Authors
Joel Poder, Ryan Brown, Andrew Howie, Johnson Yuen, Anna Ralston, Kristine Schreiber, Andrej Bece, Joseph Bucci,