Article ID Journal Published Year Pages File Type
8962852 Annals of Vascular Surgery 2018 7 Pages PDF
Abstract
The Nellix device uses polymer-filled endobags to stabilize the abdominal aortic aneurysm (AAA) sac which is described as endovascular aneurysm sealing (EVAS). We analyzed cardiac-gated computed tomography angiography scans of repaired AAA with EVAS in 4 patients to evaluate the geometry and cardiac pulsatility-induced deformation. Graft translation and aortic curvature changes were found to be minimal during the cardiac cycle. The mean ± standard deviation changes in renal-aorta angles (1.0 ± 0.9°) were less than the changes in the superior mesenteric artery-aorta angle (4.0 ± 2.1°) (P < 0.01), during the cardiac cycle, demonstrating greater stabilization of the visceral branches closer to the device. These findings confirm stabilization of the abdominal aorta during the cardiac cycle using EVAS.
Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , ,