Article ID Journal Published Year Pages File Type
9343963 Journal of Cataract & Refractive Surgery 2005 6 Pages PDF
Abstract
The quality of the axial length readings of the IOLMaster was influenced by the SNR value. However, with proper SNR evaluation and recalibration of the PCI measurements, it is possible to achieve consistent PCI readings with little variation between preoperative and postoperative measurements. These results are promising for a higher accuracy of the IOL power calculation.
Related Topics
Health Sciences Medicine and Dentistry Ophthalmology
Authors
, ,