Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
9350361 | Clinical Biomechanics | 2005 | 8 Pages |
Abstract
Interpretation. The strain increase following reaming suggests a reduction in torsional strength by over 33% which is further reduced following broaching and implant insertion. For the practicing surgeon, post-operative strength can be adversely affected by both canal preparation and implant insertion.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
Anthony M.T. Choo, Robert H. Hawkins, Brian K. Kwon, Thomas R. Oxland,