Article ID Journal Published Year Pages File Type
9460599 Journal of Aerosol Science 2005 17 Pages PDF
Abstract
The availability of a bipolar charger using soft X-ray of 9.5 keV to aerosol nanoparticle measurement was evaluated by investigating the characteristics of ion generation and by comparing the ion generation rate and the bipolar charging states of aerosol nanoparticles to those of 241Am α-ray bipolar chargers of 82 and 164μCi radiation. Ion generation of the soft X-ray charger stabilized within 7 s after being turned on and off, and remained stable for hours at a sufficient generation rate. The electrical mobilities of bipolar ions in dry nitrogen were close to each other, in contrast to those observed in the α-ray chargers. Owing to these electrical mobilities, the electrical mobility distributions of positively and negatively charged particles could be brought to almost identical states, as could the particle size distributions derived individually from the two mobility distributions. This means that inaccuracies can be minimized in measurements using a differential mobility analyzer. The effect of water vapor in ionized gas was also examined to discuss the fundamental applicability of the charger to atmospheric aerosols. For both humid nitrogen and air, the electrical mobilities of bipolar ions were found to be in accordance with those reported in the existing studies, indicating that humidity affected the electrical mobility of, in particular, negative ions.
Related Topics
Physical Sciences and Engineering Earth and Planetary Sciences Atmospheric Science
Authors
, , , ,