Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
9587315 | Journal of Magnetic Resonance | 2005 | 12 Pages |
Abstract
A novel EPR stopped-flow accessory is described which allows time-dependent cw-EPR measurements of rate constants of reactions involving paramagnetic species after rapid mixing of two liquid reagents. The EPR stopped-flow design represents a state-of-the-art, computer controlled fluid driving system, a miniresonant EPR structure with an integrated small ball mixer, and a stopping valve. The X-band EPR detection system is an improved version of that reported by Sienkiewicz et al. [Rev. Sci. Instr. 65 (1994) 68], and utilizes a resonator with two stacked ceramic dielectric rings separated by a variable spacer. The resonator with the mode TE(H)011 is tailored particularly for conditions of fast flowing and rapidly stopped aqueous solutions, and for a high time resolution. The short distance between the ball mixer and the small EPR active volume (1.8 μl) yields a measured dead time of 330 μs. A compact assembly of all parts results in minimization of disturbing microphonics. The computer controlled driving system from BioLogic with two independent stepping motors was optimized for EPR stopped-flow with a hard-stop valve. Performance tests on the EPR spectrometer ESP 300E from BRUKER using redox reactions of nitroxide radicals revealed the EPR stopped-flow accessory as an advanced, versatile, and reliable instrument with high reproducibility.
Related Topics
Physical Sciences and Engineering
Chemistry
Physical and Theoretical Chemistry
Authors
Günter Lassmann, Peter Paul Schmidt, Wolfgang Lubitz,