Abstract
Several factors should be considered if a commercial OF collection device is used. In vitro collection experiments demonstrated that the mean collection volume varied between devices from 0.82 to 1.86Â mL. The percentage of the collected volume that could be recovered from the device varied from 18% to 83%. In vitro experiments demonstrated considerable variation in the recovery of amphetamines (16-59%), opiates (33-50%), cocaine and benzoylecgonine (61-97%), carboxy-THC (0-53%) and PCP (9-56%). Less variation in collection volume, volume recovered and drug recovery was observed intra-device. The THC stability was evaluated in a common commercial collection protocol. Samples in the collection buffer were relatively stable for 6 weeks when stored frozen. However, stability was marginal under refrigerated conditions and poor at room temperature. Very little has been published on the efficacy of using IgG concentration, or any other endogenous marker, as a measure of OF specimen validity. Preliminary rinsing experiments with moderate (50Â mL and 2Â ÃÂ 50Â mL) volumes of water did not reduce the OF IgG concentration below proposed specimen validity criteria. In summary, obvious and more subtle variables in the OF collection may have pronounced effects on OF-drug concentrations. This has rarely been acknowledged in the literature, but should to be considered in OF drug testing, interpretation of OF-drug results and future research studies.