Abstract
Besides the accuracy and the precision of the measurements of the data points, several important parameters affect the accuracy of the adsorption isotherms that are derived from the data acquired by frontal analysis (FA). The influence of these parameters is discussed. First, the effects of the width of the concentration range within which the adsorption data are measured and of the distribution of the data points in this range are investigated. Systematic elimination of parts of the data points before the calculation of the nonlinear regression of the data to the model illustrates the importance of the numbers of data points (1) within the linear range and (2) at high concentrations. The influence of the inaccuracy of the estimate of the column hold-up volume on each adsorption data point, on the selection of the isotherm model, and on the best estimates of the adsorption isotherm parameters is also stressed. Depending on the method used to measure it, the hold-up time can vary by more than 10%. The high concentration part of the adsorption isotherm is particularly sensitive to errors made on t0,exp and as a result, when the isotherm follows bi-Langmuir isotherm behavior, the equilibrium constant of the low-energy sites may change by a factor 2. This study shows that the agreement between calculated and experimental overloaded band profiles is a necessary condition to validate the choice of an adsorption model and the calculation of its numerical parameters but that this condition is not sufficient.