Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
9748858 | Journal of Chromatography A | 2005 | 9 Pages |
Abstract
A fast determination of isoflavones (genistin, genistein, daidzein, daidzin, glycitin, glycitein, ononin, formononetin, sissotrin and biochanin A) by HPLC/UV-vis-DAD working at 254 nm is described. An Atlantis dC18 fast reversed-phase chromatographic column (20 mm Ã 2.1 mm, 3 μm particle size) was used at a flow rate 0.35 ml minâ1 of a mobile phase consisted from 0.1% (v/v) acetic acid (A) at pH 3.75 and methanol. (B). A linear gradient profile was used for separation at the column temperature 36 °C. Limits of detection (LODs for 3 S/N criterion) per sample injection (5 μl) ranged from 166.2 to 17.0 fmol (9.4-1.1 ng mlâ1) for biochanin A and genistin, respectively. The recoveries 96-106% were obtained for the different concentrations of the isoflavones (RSDs 2-8%). The pressurized liquid extraction/HPLC/UV-vis-DAD method was used for the determination of the isoflavones in soy bits (28-962 μg gâ1 dry weight). The proposed procedure is faster (ca. 8 min) without loosing its separation efficiency (up to 10 isoflavonoids can be determined) and sensitivity (tens to hundreds fmol).
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
BoÅivoj Klejdus, Radka Mikelová, Jitka Petrlová, David PotÄÅ¡il, VojtÄch Adam, Marie Stiborová, Petr Hodek, Jan Vacek, René Kizek, Vlastimil KubáÅ,