Article ID Journal Published Year Pages File Type
9748963 Journal of Chromatography A 2005 8 Pages PDF
Abstract
A method based on the presence of a dynamic pH junction within the capillary to induce band narrowing for enhanced detection sensitivity for some peptides is presented. This technique is predicated on a sharp reduction in an analyte's migration velocity following a reversal of its electrophoretic direction from the acidic sample zone to the basic BGS zone. Larger-than-usual injection volumes of samples in relatively high-conductivity matrices were enabled, without degrading peak shape, resolution and efficiency. The size of the original sample plug was reduced by as much as 38-fold, and improvement in detector response in terms of peak height by as much as 124-fold was obtained. The effects of pH and concentration of the sample matrix, and the length of sample injection on the efficiency of the technique are discussed.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,