Abstract
The capability of hemimicelles-based solid phase extraction (SPE)/liquid chromatography/atmospheric pressure chemical ionisation in positive mode, ion trap mass spectrometry (LC/(APCI+-IT)-MS) for the concentration, separation and quantitation of non-ionic surfactants has been investigated. Concentration was based on the formation of mixed aggregates of analytes [alkylphenol ethoxylates (APE, octyl and nonyl) and alkyl ethoxylates (AE, C12-C16)] with the anionic surfactant sodium dodecyl sulphate (SDS) that is adsorbed on alumina. Parameters affecting SPE were investigated on the basis that hemimicelles are dynamic entities in equilibrium with the aqueous phase. The performance of ion trap mass spectrometry for MS and MS/MS quantitation of non-ionic homologues was assessed. Recoveries of analytes from wastewater influent and effluent and river water samples ranged between 91 and 98% and were found independent on the length of the alkyl chain under the optimised conditions. Anionic surfactants did not interfere to the levels found in environmental samples. The detection limits ranged between 14 and 111 ng/l for wastewater influent, 10 and 40 for wastewater effluent and 4 and 35 for river water, after concentration of 250, 500 and 750 ml of sample, respectively. The approach was applied to the determination of AE and APE in influent and effluent samples from four wastewater treatment plants and four river samples. The concentrations of individual non-ionic surfactants found ranged between 0.3 and 373 μg/l.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,