Article ID Journal Published Year Pages File Type
9753276 Journal of Chromatography B 2005 7 Pages PDF
Abstract
Monitoring in vivo oxidative stress implicates the evaluation of damage and defence parameters by well-established, validated methods. We report two optimized and validated HPLC methods for measurement of malondialdehyde (MDA) and fat-soluble vitamins in rat plasma. For the MDA method, optimization experiments of the thiobarbituric acid test resulted in the addition of 1% butylhydroxytoluene to the reaction mixture and in a heating time reduction to 40 min, ensuring inhibition of further lipid peroxidation during the test. Validation experiments showed good linearity, precision and recovery. The use of HPLC with coulometric array detection technology permits simultaneous and sensitive analysis of different fat-soluble vitamins and related compounds (tocopherols, retinoids, carotenoids and coenzyme Q10), which are identified by both retention time and electrochemical characteristics. Furthermore, this method is extended to the analysis of coenzyme Q9, the predominant homologue in rats. Validation experiments with rat plasma gave good results.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,