A high-performance liquid chromatographic (HPLC) method with fluorimetric detection was developed for the simultaneous determination of ibogaine and noribogaine in human plasma using fluorescein as internal standard. This method involved a solid phase extraction of the compounds from plasma using N-vinylpyrrolidone-divinybenzene copolymer cartridges. Separation of the three analytes was performed on a reversed-phase Supelcosil C18 analytical column (75 mm Ã 4.6 mm i.d., 3 μm particle size). The excitation wavelength was set at 230 nm for the first 15.8 min and then at 440 nm for the following 14.2 min; the emission wavelength was set at 336 nm for the first 15.8 min and then at 514 nm for the following 14.2 min. Obtained from the method validation, inter-assay precision was 6.0-12.5% and accuracy was 95.4-104%. The extraction efficiencies of the assay were higher than 94% and were constant across the calibration range. The lower limits of quantitation were 0.89 ng/ml for ibogaine and 1 ng/ml for noribogaine; at these levels, precision was â¤17% and accuracy was 95-105%. In this paper, extensive stability testing was undertaken using a wide range of storage conditions. Special attention must be paid to sample handling to avoid light degradation of the compounds.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Violeta KontrimaviÄiÅ«tÄ, Michel Larroque, Vitalis Briedis, Delphine Margout, Françoise Bressolle,