Article ID Journal Published Year Pages File Type
9754410 Journal of Pharmaceutical and Biomedical Analysis 2005 6 Pages PDF
Abstract
The performance of four commercially available polysaccharide-based chiral stationary phases, Chiralcel OD, Chiralcel OJ, Chiralpak AD and Chiralpak AS was evaluated after several cycles of extended multimodal operations. Acetonitrile, methanol, ethanol and their mixtures were chosen for polar-organic mobile phases, ethanol-n-heptane mixture and n-heptane were selected for normal-phase mode. Retention factor (k), selectivity (α), resolution (Rs) and theoretical plate count (N) were the chosen parameters to describe the column performance. One racemate for which all four columns have shown enantioselectivity was chosen as test compound. After 15 cycles of multimodal operations a slight decrease was seen in the retention factors (k) however, column efficiency, selectivity (α) and resolution (Rs) were maintained.
Keywords
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,