Abstract
A flow injection square wave cathodic stripping voltammetric method has been developed for the determination of sertraline in a pharmaceutical preparation. The method shows linearity between peak current intensity and sertraline concentration for the interval between 0.20 Ã 10â6 and 1.20 Ã 10â6 mol Lâ1. Limits of detection and quantification were found to be 1.5 Ã 10â7 and 5.0 Ã 10â7 mol Lâ1, respectively. Up to 70 samples per hour can be analysed with a good precision (R.S.D. = 2.5%). The proposed method was successfully applied to the determination of sertraline in a commercial product. In the voltammetric determination of sertraline in flow, a high sample rate is obtained at reduced costs, opening the possibility to compete with the chromatographic methods generally used for this analysis.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Henri P.A. Nouws, Cristina Delerue-Matos, Aquiles A. Barros, José A. Rodrigues,