Article ID Journal Published Year Pages File Type
9754517 Journal of Pharmaceutical and Biomedical Analysis 2005 4 Pages PDF
Abstract
A flow injection square wave cathodic stripping voltammetric method has been developed for the determination of sertraline in a pharmaceutical preparation. The method shows linearity between peak current intensity and sertraline concentration for the interval between 0.20 × 10−6 and 1.20 × 10−6 mol L−1. Limits of detection and quantification were found to be 1.5 × 10−7 and 5.0 × 10−7 mol L−1, respectively. Up to 70 samples per hour can be analysed with a good precision (R.S.D. = 2.5%). The proposed method was successfully applied to the determination of sertraline in a commercial product. In the voltammetric determination of sertraline in flow, a high sample rate is obtained at reduced costs, opening the possibility to compete with the chromatographic methods generally used for this analysis.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,