Article ID Journal Published Year Pages File Type
9844921 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2005 8 Pages PDF
Abstract
The response function of a high-resolution high-energy photon spectrometer (HHS), consisting of a twin Ge detector and BGO (Bi3Ge4O12) anti-coincidence shields, has been measured for 10.763 MeV monochromatic γ-rays generated in a resonant reaction 27Al(p,γ)28Si. The experimental response functions measured with and without the anti-coincidence were compared with results by Monte Carlo simulations, in which a pulse-height degradation effect associated with a surface channel of the twin Ge detector has been included. Excellent agreement was obtained between the experimental response functions and the simulations.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , ,