Article ID Journal Published Year Pages File Type
9845300 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2005 10 Pages PDF
Abstract
When using two detectors to measure the emission of gamma rays from a bulky sample, the count rates ratio of the detectors can be used to locate a radioactive source in a bulky volume. This process may be needed for locating a source in a scrap or waste container. Simple functions for location determination can be developed for the non-absorbing and pure-absorbing media, but for the general case, a more sophisticated method must be used. A simple and practical algorithm was developed, which can predict the location of a point source in a bulk volume quite accurately, with an average possible deviation of up to about 1 cm, except for extreme combinations of dimension and absorption coefficient, but also in this case the maximal expected deviations are not more than several centimeters. The algorithm was checked experimentally for several bulk volumes with length of up to about 50 cm and with different length/absorption coefficient combinations. The experimental validation was performed for a great range of absorption coefficients (from μ=0.0001 to 0.2 cm−1), covering the practical range of waste containers compositions.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , ,