Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
9845320 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2005 | 4 Pages |
Abstract
We report the performances of a 500 μm pixellated silicon sensor bonded to the photon counting chip Medipix2 [1]. In order to perform an absolute characterization of our detector, we measured both the pre-sampling MTF and NPS with respect to the International standard IEC-62220-1. From those data we have been able to extract the Detective Quantum Efficiency (DQE) and hence to assess the suitability of our detector for X-ray medical imaging purpose. Due to poor absorption of the Si at 70 kV the DQE peaked at 0.06 for null frequency. Nevertheless, these results are very promising since thicker Si or more absorbing material such as GaAs will soon be available.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Benjamin Surre, Mario Caria, Julien Chaput, Thierry Hassoun, Fabien Laverroux, Laurent Sarry,