Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
9845322 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2005 | 4 Pages |
Abstract
A new type of experimental technique to investigate the depth-of-interaction (DOI) dependence in small scintillator elements designed for high-resolution animal PET is tested. An Oxford-type nuclear microprobe was used to irradiate a lutetium oxyorthosilicate (LSO) crystal (2Ã2Ã10 mm3) with a highly focused, (few micrometers diam.), 2 MeV He+ beam. Pulse height spectra from a PMT attached to one end of the LSO crystal were collected in list mode. By scanning the beam over the crystal high-resolution 2D-images are generated and thus the DOI effect in the scintillator could be investigated. The obtained results of the overall DOI dependence confirm previous findings using γ-rays. Since the present experimental setup allows obtaining data with several orders of magnitude better spatial resolution than with collimated γ-beam, the method can be applied for variety of problems where high spatial resolution is required.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
A. Simon, L. Balkay, G. Kalinka, A. Kerek, D. Novák, A. Sipos, J. Végh, L. Trón, J. Molnár,