Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
9845758 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2005 | 8 Pages |
Abstract
This report describes a method to effectively discriminate between tritium and other interfering airborne radioactive gases using ionization cluster size information from a double gas electron multiplier (DGEM) and a suitable readout array. The readout array consists of 121 pads (an 11Ã11 matrix of 4Ã4 mm pads on 6 mm centres) on a printed circuit board, each connected to a charge sensitive amplifier. The prototype-automated discriminating monitor typically indicated a spillover fraction, in the tritium channel, in the order of ±5% for binary mixtures of tritium and 14C, 85Kr, 133Xe or 41Ar. A custom-made dynamic cluster-fitting algorithm gives results in real-time with very good accuracy under extreme conditions.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
R.A. Surette, J. Dubeau,