9845802 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2005 15 Pages PDF
Abstract
Detector telescopes having a ΔE-E configuration, consisting of either a Si surface barrier detector or a Si strip detector followed by a NaI(Tl) stopping detector, are described for measuring intermediate mass fragments in 12C-induced reactions to the continuum. These include the stable break-up fragments 7Be and 9Be, as well as the resonant fragment 8Be. Energy calibration of the stopping detectors is based on a scintillation light-response model. Empirical approximating functions have been selected for inverting the theoretical functions relating light output to particle energy, in order to achieve calibration during rapid data sorting. The detector efficiency for resonant fragments is obtained by means of a Monte Carlo simulation. A dedicated eight-channel preamplifier module with good operational stability in vacuum, specifically designed for Si-strip detectors, is also described.