Article ID Journal Published Year Pages File Type
103474 Legal Medicine 2015 4 Pages PDF
Abstract

•The efficacy of a new on-site immunoassay drug-screening device Triage-TOX was examined.•This device was applicable to drug screening using postmortem urine and pericardial fluid.•Triage-TOX showed high sensitivity for benzodiazepines, with objective data presentation.•However, its combination with another device can help minimize misinterpretation.•Provision of systematic instrumental identification and quantification is essential.

Instrumental identification of drugs with quantification is essential in forensic toxicology, while on-site immunoassay urinalysis drug-screening devices conveniently provide preliminary information when adequately used. However, suitable or sufficient urine specimens are not always available. The present study evaluated the efficacy of a new on-site immunoassay drug-screening device Triage-TOX (Alere Inc., San Diego, CA, USA), which has recently been developed to provide objective data on the one-step automated processor, using 51 urine and 19 pericardial fluid samples from 66 forensic autopsy cases, compared with Triage-Drug of Abuse (DOA) and Monitect-9. For benzodiazepines, the positive predictive value and specificity of Triage-TOX were higher than those of Triage-DOA; however, sensitivity was higher with Monitect-9, despite frequent false-positives. The results for the other drugs with the three devices also included a few false-negatives and false-positives. These observations indicate the applicability of Triage-TOX in preliminary drug screening using urine or alternative materials in routine forensic autopsy, when a possible false-negative is considered, especially for benzodiazepines, providing objective information; however, the combined use of another device such as Monitect-9 can help minimize misinterpretation prior to instrumental analysis.

Graphical abstractThe present study evaluated the efficacy of a new on-site immunoassay drug-screening device Triage-TOX (Alere Inc., San Diego, CA, USA), which has recently been developed to provide objective data on the one-step automated processor, using 51 urine and 19 pericardial fluid samples from 66 forensic autopsy cases, compared with Triage-Drug of Abuse (DOA) and Monitect-9.Figure optionsDownload full-size imageDownload high-quality image (144 K)Download as PowerPoint slide

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,