Abstract
There is relatively little guidance on the situated ergonomic assessment of medical devices, and few case studies that detail this type of evaluation. This paper reports results of a detailed case study that focuses on the design and use of a modern blood glucose meter on an oncology ward. We spent approximately 150Â h in-situ, over 11 days and 4 nights, performing observations and interviews with users. This was complemented by interviews with two staff with oversight and management responsibility related to the device. We identified 19 issues with the design and use of this device. These issues were grouped into 7 themes which can help guide the situated study of medical devices: usability, knowledge gaps and mental models, workarounds, wider tasks and equipment, the patient, connection between services, and policy.
Keywords
Related Topics
Physical Sciences and Engineering
Computer Science
Human-Computer Interaction
Authors
Dominic Furniss, Paolo Masci, Paul Curzon, Astrid Mayer, Ann Blandford,