Article ID Journal Published Year Pages File Type
10529080 Analytica Chimica Acta 2005 6 Pages PDF
Abstract
The use of hydride generation coupled with axial view inductively couple plasma atomic emission spectrometry was presented for the determination of selenium in plant samples. The chemical factors affecting potentially the hydride generation efficiency (hydrochloric acid, sodium borohydride and sodium hydroxide concentrations) were assessed through investigation of chemical interference, accuracy and repeatability. The accuracy of measurements was not affected by elements present in high concentration in the plant matrix (K, Ca, Mg, and P). No interference was also observed with transition metals. Using a real sample (maize) with standard additions, decreases of recoveries were sometimes observed for 0.1% (m/v) NaOH, and attained 13.8% in the most unfavourable case. The final accuracy of the method was verified by using two certified reference materials: CRM 402 (white clover) and CRM 279 (sea lettuce). No statistically significant differences were obtained between the measured concentrations and the certified values. The optimized method was found sensitive (detection limit 0.15 μg l−1), reliable and repeatable (R.S.D. between 1.3% and 4.0%).
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,