Article ID Journal Published Year Pages File Type
108025 Digital Applications in Archaeology and Cultural Heritage 2015 6 Pages PDF
Abstract

•The MicroScribe is an accurate and precise instrument (errors <1 mm).•The accuracy is maintained across the workspace.•Both linear distance and geometric morphometry are accurate.•The accuracy is maintained with different positions of the digitizer arm.

Portable three-dimensional digitizers facilitate the “digitize” and “compare” operational areas of Virtual Anthropology. However, these measurement devices must be adequately accurate and precise for their intended purpose. We determined that varying the arm configurations led to standard deviations less than 0.23 mm. The distances between pairs of points among 10 points on a reference object at 16 locations within the workspace had an average error of 0.239 mm; only three of the 720 measures had an error greater than 1 mm. Through assessing geometric morphometry, the average landmark standard deviation of the 10 points at 16 locations across the workspace was only 0.12 mm. The different areas of the workspace and different arm joint configurations do not meaningfully influence the measurement precision and accuracy. Therefore, the MicroScribe G2X has suitable accuracy and precision to be an appropriate research tool for many anthropological, forensic and biomechanical studies.

Graphical abstractMicroScribe digitizing arm reaching to points on the reference objectFigure optionsDownload full-size imageDownload as PowerPoint slide