Article ID Journal Published Year Pages File Type
1162693 Analytica Chimica Acta 2016 8 Pages PDF
Abstract

•The enhanced electrochemiluminescence of luminol by AuNPs/TiNTs.•An ECL biosensor for HbA1c assay with ultra-high sensitivity.•A promising disposable device for diabetic diagnosis and treatment even for POCT.•The excellent regression of detected results with gold-standard method.

A glycated hemoglobin (HbA1c) biosensor with high performance has been constructed in this work. Here the fructosyl amino acid oxidase was immobilized onto a pre-functionalized indium tin oxide glass with titania nanotubes decorated with gold nanoparticles. The property of nanocomposite was characterized by transmission electromicroscopy, scanning electron microscopy, electrochemistry and spectroscopy. Under the optimum conditions, fructosyl valine was detected by this biosensor. It exhibited a linear detection range from 4.0 × 10−9 M to 7.2 × 10−7 M, and a limit of detection for 3.8 × 10−9 M at the signal-to-noise ratio of 3. Thus the HbA1c level in whole blood samples of healthy individuals or diabetic patients were evaluated with designed biosensor after pre-treatment of hydrolysis. The results of our detection were closely consistent with that of the standard method. At the same time, our biosensor has some advantages including high sensitivity, disposable usage and low cost, which implies its great promising application in point-of-care testing of HbA1c.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slide

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,