Article ID Journal Published Year Pages File Type
1163054 Analytica Chimica Acta 2016 9 Pages PDF
Abstract

•μ-SPE coupled with IDMS for the measurement of carbamazepine.•The method is the first report of coupling μ-SPE with IDMS.•μ-SPE is fast, time- and cost-effective, and environmentally friendly.•Highly precise and accurate IDMS measurement of carbamazepine in surface water.•The method facilitates the fast and accurate monitoring of anthropogenic markers.

To achieve fast and accurate analysis of carbamazepine in surface water, we developed a novel porous membrane-protected micro-solid-phase extraction (μ-SPE) method, followed by liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) analysis. The μ-SPE device (∼0.8 × 1 cm) was fabricated by heat-sealing edges of a polypropylene membrane sheet to devise a bag enclosing the sorbent. The analytes (both carbamazepine and isotope-labelled carbamazepine) were first extracted by μ-SPE device in the sample (10 mL) via agitation, then desorbed in an organic solvent (1 mL) via ultrasonication. Several parameters such as organic solvent for pre-conditioning of μ-SPE device, amount of sorbent, adsorption time, and desorption solvent and time were investigated to optimize the μ-SPE efficiency. The optimized method has limits of detection and quantitation estimated to be 0.5 ng L−1 and 1.6 ng L−1, respectively. Surface water samples spiked with different amounts of carbamazepine (close to 20, 500, and 1600 ng L−1, respectively) were analysed for the validation of method precision and accuracy. Good precision was obtained as demonstrated by relative standard deviations of 0.7% for the samples with concentrations of 500 and 1600 ng kg−1, and 5.8% for the sample with concentration of 20 ng kg−1. Good accuracy was also demonstrated by the relative recoveries in the range of 96.7%–103.5% for all samples with uncertainties of 1.1%–5.4%. Owing to the same chemical properties of carbamazepine and isotope-labelled carbamazepine, the isotope ratio in the μ-SPE procedure was accurately controlled. The use of μ-SPE coupled with IDMS analysis significantly facilitated the fast and accurate measurement of carbamazepine in surface water.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slide

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,