Article ID Journal Published Year Pages File Type
1163347 Analytica Chimica Acta 2015 7 Pages PDF
Abstract

•Life support system for human spaceflight missions.•On-line chemical sensing in water recycling processes.•Potentiometric determination of ammonium ion using a gas diffusion step.•Samples of the Concordia station (Antarctica) gray water treatment plant.

The design, construction and evaluation of a versatile cyclic olefin copolymer (COC)-based continuous flow potentiometric microanalyzer to monitor the presence of ammonium ion in recycling water processes for future manned space missions is presented. The microsystem integrates microfluidics, a gas-diffusion module and a detection system in a single substrate. The gas-diffusion module was integrated by a hydrophobic polyvinylidene fluoride (PVDF) membrane. The potentiometric detection system is based on an all-solid state ammonium selective electrode and a screen-printed Ag/AgCl reference electrode. The analytical features provided by the analytical microsystem after the optimization process were a linear range from 0.15 to 500 mg L−1 and a detection limit of 0.07 ± 0.01 mg L−1. Nevertheless, the operational features can be easily adapted to other applications through the modification of the hydrodynamic variables of the microfluidic platform.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slide

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,