Abstract

•High-precision determination of nitrite and nitrate in seawater.•Use of quadruple isotope dilution.•Aqueous Et3O+[BF4]− derivatization chemistry for GCMS analysis of nitrite and nitrate.

Quadruple isotope dilution mass spectrometry (ID4MS) has been applied for simultaneous determination of nitrite and nitrate in seawater. ID4MS allows high-precision measurements and entails the use of isotopic internal standards (18O-nitrite and 15N-nitrate). We include a tutorial on ID4MS outlining optimal experimental design which generates results with low uncertainties and obviates the need for direct (separate) evaluation of the procedural blank. Nitrite and nitrate detection was achieved using a headspace GCMS procedure based on single-step aqueous derivatization with triethyloxonium tetrafluoroborate at room temperature. In this paper the sample preparation was revised and fundamental aspects of this chemistry are presented. The proposed method has detection limits in the low parts-per-billion for both analytes, is reliable, precise, and has been validated using a seawater certified reference material (MOOS-2). Simplicity of the experimental design, low detection limits, and the use of quadruple isotope dilution makes the present method superior to the state-of-the-art for determination of nitrite and nitrate, and an ideal candidate for reference measurements of these analytes in seawater.