1164898 | Analytica Chimica Acta | 2013 | 7 Pages |
A series of Boron-Doped Diamond (BDD) ultramicroelectrode arrays were fabricated and investigated for their performance as electrochemical sensors to detect trace level metals such as cadmium. The steady-state diffusion behavior of these sensors was validated using cyclic voltammetry followed by electrochemical detection of cadmium in water and in human urine to demonstrate high sensitivity (>200 μA ppb−1 cm−2) and low background current (<4 nA). When an array of ultramicroelectrodes was positioned with optimal spacing, these BDD sensors showed a sigmoidal diffusion behavior. They also demonstrated high accuracy with linear dose dependence for quantification of cadmium in a certified reference river water sample from the U.S. National Institute of Standards and Technology (NIST) as well as in a human urine sample spiked with 0.25–1 ppb cadmium.
Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slideHighlights► An electrochemical sensor capable of detecting cadmium at parts-per-billion levels in urine. ► A novel fabrication method for Boron-Doped Diamond (BDD) ultramicroelectrode (UME) arrays. ► Unique combination of BDD UME arrays and a differential pulse voltammetry algorithm. ► High sensitivity, high reproducibility, and very low noise levels. ► Opportunity for portable operation to assess on-site personal exposure.