Article ID Journal Published Year Pages File Type
1165870 Analytica Chimica Acta 2012 4 Pages PDF
Abstract

Two quantification strategies for one-point standard addition calibration have been compared mathematically. One strategy involved the extrapolation of measurement points to their intercept with the x-axis to determine the analyte content in the unknown sample, and the other strategy is based upon direct calculation of the analyte content in the unknown sample using the instrumental responses obtained during measurement. The cases of both conventional standard addition calibration (C-SAC) and sequential standard addition calibration (S-SAC) have been considered. The homoscedastic situation has been considered, where the absolute precision of the instrumental response is constant. It has been determined that the precision ratio of the two strategies is dependent on surprisingly simple parameters: such as the sample to standard mass ratio (for C-SAC) and the analyte content ratio (for S-SAC).

Graphical abstractThe precision characteristics of two quantification methods for one point standard addition calibration have been quantified and compared.Figure optionsDownload full-size imageDownload as PowerPoint slideHighlight► When using one-point standard addition calibration two quantification strategies are possible. ► This work has compared the precision characteristics of these two methods: extrapolation and use of analytical intensities. ► It has been determined that the precision ratio of the two strategies is dependent on surprisingly simple parameters. ► For conventional standard addition the relevant parameter is the sample to standard mass ratio. ► For sequential standard addition the relevant parameter is the sample to standard analyte content ratio.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,