Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1168009 | Analytica Chimica Acta | 2010 | 8 Pages |
Abstract
A new miniaturization approach to create micro- and nanoscale ion selective electrodes (ISEs) was demonstrated and the concept tested with an environmentally relevant chromate-selective membrane consisting of 7.7:62.2:31.1 wt % Aliquat336:2-NPOE:PVC. Apertures of 100 nM and 30 μM dimensions were made using MEMS fabrication techniques and functionalized through a macroscale application of solvent polymeric membrane. Performance studies for the microscale ISE showed a response slope of â58.6 ± 5.6 mV decadeâ1 and limit of detection (LOD) of 2.1 Ã 10â5 ± 1.1 Ã 10â5 M, versus â65.2 ± 4.2 mV decadeâ1 and 1.8 Ã 10â5 ± 6 Ã 10â6 M for the nanoscale ISE. This was consistent with control studies with carefully conditioned coated wire electrodes, which demonstrated a response slope of â61.7 ± 2.4 mV decadeâ1 and a LOD of 3.0 Ã 10â6 ± 1 Ã 10â6 M. Response times for the best micro- and nanoscale ISEs were in the 10-20 s timeframe. Electrical resistance measurements were in the GΩ range for the microscale ISEs and nanoscale ISEs. Appropriate ISE geometry was confirmed through AFM measurements and calculations based on electrical properties for micro- and nanoscale apertures. These micro- and nanoscale ISEs are expected to have significant impact in the field of microscale analytical processes.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Daniel C. Rieck, Bingwen Liu, Bong-Jae Park, David F. Moffett, David A. Kidwell, Gary J. Cheng, Bernard J. Van Wie,