Article ID Journal Published Year Pages File Type
1168069 Analytica Chimica Acta 2010 5 Pages PDF
Abstract

A novel capillary electrophoresis and amperometric detection method was achieved by adding an electroactive additive (3,4-dihydroxybenzylamine, 3,4-DHBA) to the running buffer, so that both electroactive and non-electroactive food preservatives were simultaneously determined. Under the selected optimum conditions, four electroactive preservatives (methylparaben, ethylparaben, propylparaben and butylparaben) and two non-electroactive preservatives (potassium sorbate and sodium lactate) were well separated and sensitively detected with detection limits (S/N = 3) ranging from 1.06 × 10−8 to 2.73 × 10−6 g mL−1. This method has been successfully employed for the determination of both electroactive and non-electroactive preservatives in several food commodities.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,