Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1168187 | Analytica Chimica Acta | 2009 | 7 Pages |
Abstract
A novel electrochemical methods namely standard free anodic stripping voltammetry and anodic stripping voltammetric titration are proposed for determination of dissolved sulfide concentration. 2Ag+ + S2− → Ag2S reaction is used to provide the information. The anodic stripping voltammetric response of unreacted silver-ions at the glassy carbon electrode is used as analytical signal. Results reliability and accuracy are confirmed by analysis of model solutions, spiked natural and tap waters and recovery study, with a recovery of 100 ± 5% (n = 7) obtained. The approaches show the detection limit (3σblank) of 2–5 × 10−10 mol L−1 and the relative standard deviation of 2–5% for repeated measurements.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
T.V. Titova, N.S. Borisova, N.F. Zakharchuk,