Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1169454 | Analytica Chimica Acta | 2008 | 8 Pages |
Abstract
A method based on ultra-performance liquid chromatography-electrospray ionisation-tandem mass spectrometry for the determination of six thyreostatic drugs in thyroid tissue has been optimised and validated in accordance with the Decision 2002/657/EC. Samples are extracted with methanol and the extracts cleaned-up on silica cartridges. The recoveries range from 40% for 6-phenyl-2-thiouracil to 79% for 2-thiouracil. Quantification is carried out with blank tissue samples spiked with the analytes in the range 25-500 μg kgâ1. 5,6-Dimethyl-2-thiouracil is used as internal standard. CCα and CCβ are in the ranges 4.3-16.1 μg kgâ1 and 8.7-20.7 μg kgâ1, respectively. Accuracy, expressed as percentage of error, is lower than 6% and relative standard deviation in reproducibility conditions falls between 5.6 and 10.3%. Nowadays, the proposed method is routinely implemented in the laboratory of the Agència de Salut Pública de Barcelona and allows processing of up to 20 samples per day.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
S. AbuÃn, F. Centrich, A. Rúbies, R. Companyó, M.D. Prat,