Abstract

A fast and highly sensitive high performance liquid chromatographic/ion-trap mass spectrometric method (LC/MS) has been developed for simultaneous determination of ethynylestradiol (EE2), gestodene (GES), levonorgestrel (LNG), cyproterone acetate (CPA) and desogestrel (DES). Among three types of sorbents tested (C8, C18 and phenyl) from two suppliers, the best separation was achieved on reverse phase Zorbax SB-Phenyl column using aqueous methanol as a mobile phase. A linear gradient profile from 70 up to 100% (v/v) in 7th min, kept constant at 100% up to 10th min and followed by a negative gradient to 70% of methanol up to 12th min was used for elution. Applicability of electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI) and influence of the mobile phase composition, its flow rate, capillary/vaporizer temperature of API source and in-source fragmentor voltage ionization are discussed. The on-column limits of quantification (10 S/N) were 300 pg of EE2, 14 pg of GES and LNG, 4 pg of CPA and 960 pg of DES per injection (1 μL) using APCI with data collection in selected ion monitoring (SIM) mode. The analytical performance of the method was evaluated using the determination of EE2, GES, LNG, CPA and DES in contraceptives and river water samples.