Article ID Journal Published Year Pages File Type
1171291 Analytica Chimica Acta 2006 8 Pages PDF
Abstract

In this paper, an ordered mesoporous alumina coating was prepared and applied to capillary microextraction (CME) of trace Co, Ni and Cd for the first time. The coated capillary was used for on-line coupling CME with inductively plasma mass spectrometry (ICP-MS) for the determination of trace metals of Co, Ni and Cd. The porous structure of Al2O3 coating was examined by SEM and TEM. The effects of the extraction parameters including pH, sample flow rate and volume, elution solution and interfering ions on the recovery of analytes have been investigated and optimized. Under the optimized conditions, the limits of detection were 0.33, 1.5 and 1.4 ng L−1 for Co, Ni and Cd, respectively, with a preconcentration factor of 10 times. The precisions for all investigated elements were 2.7, 4.1 and 2.5% (c = 0.05 ng L−1, n = 7), for Co, Ni and Cd, respectively, and the sample frequency was 8 h−1.The proposed method was successfully applied for the analysis of trace metals in water, rice and urine samples with the recovery of 94–105%. In order to validate the proposed method, two certified reference materials of GBW 0913 human urine and NIES No.10-b rice flour were analyzed, and the determination values are in good agreement with the certified values. The ordered mesoporous Al2O3 coated capillary can be used more than 20 times without decreasing the extraction efficiency.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,