Article ID Journal Published Year Pages File Type
1171301 Analytica Chimica Acta 2006 8 Pages PDF
Abstract

This work reports a sequential-injection analysis (SIA) method for the enzymatic assay of glucose with soluble glucose oxidase (GOD) and on-line sample dilution with chemiluminescence (CL) detection. A zone of sample was aspirated in the holding coil of the SIA manifold and, if necessary, was diluted on-line by means of an auxiliary dilution conduit. Then, a zone of GOD was aspirated adjacent to the sample zone and a stopped-flow period was applied to allow the enzymatic reaction to proceed with production of hydrogen peroxide. Then, zones of a catalyst (Co(II) solution) and alkaline luminol were aspirated into the holding coil. Finally, the flow was reversed and the stacked zones were sent to a flow-cell located in front of a photomultiplier tube (PMT) that monitored the CL intensity. The linear dynamic range was 1 × 10−5–1 × 10−3 mol L−1 glucose, the coefficient of variation at 8 × 10−5 mol L−1 of glucose was sr = 3.1% (n = 8), the limit of detection at the 3σ level was cL = 1 × 10−6 mol L−1 and the sampling frequency was 28 h−1. With on-line dilution by a factor of 1/200, the linear range could be extended up to 0.2 mol L−1 glucose. The advantages of the proposed method are the simple manifold and instrumentation used, the scope for automated on-line dilution, the low consumption of sample and reagents and the elimination of enzyme immobilisation procedures. The method was applied to the analysis of commercial drinks and honey with percent relative errors in glucose determination in the range 100 ± 6.1%.