Abstract
A flow-injection analytical (FIA) system, comprised of an electrochemical detector with a fructosyl-peptide oxidase (FPOX-CET) reactor and a flow-type spectrophotometer, was proposed for the simultaneous measurement of glycohemoglobin and total hemoglobin in blood cell. The blood cell samples were hemolyzed with a surfactant and then treated with protease. In the first stage of operation, total hemoglobin in digested sample was determined spectrophotometrically. In the second stage, fructosyl valyl histidine (FVH) released from glycohemoglobin by the selective proteolysis was determined specifically using the electrochemical detector with the FPOX-CET reactor. The FIA system could be automatically processed at an analytical speed of 40 samples per hour. The proposed assay method could determine selectively only the glycated N-terminal residue of β-chain in glycohemoglobin and total hemoglobin in blood cell. The enzymatic hemoglobin A1c (HbA1c) value calculated by the concentration ratio of the FVH to total hemoglobin, was closely correlated with the HbA1c values certified by the Japan Diabetic Society (JDS) and the International Federation of Clinical Chemistry (IFCC).