Abstract

An optical bio-sniffer for methyl mercaptan (MM) one of major odorous chemicals in halitosis (bad breath) was constructed by immobilizing monoamine oxidase type A (MAO-A) onto a tip of a fiber optic oxygen sensor (od: 1.59 mm) with an oxygen sensitive ruthenium organic complex (excitation: 470 nm, fluorescent: 600 nm). A flow cell for circulating buffer solution was applied to rinse and clean the tip of the device like nasal mucosa. In order to amplify the bio-sniffer output, a substrate regeneration cycle caused by coupling MAO-A with l-ascorbic acid (AsA) as reducing reaction with reagent system was applied to the sensor system. After evaluating the sensor characteristics using a gas flow measurement system with a gas generator, the optical bio-sniffer was applied to expired gases from healthy male volunteers for halitosis analysis as a physiological application.The optical bio-sniffer was applied to detect the oxygen consumption induced by MAO-A enzymatic reaction (and AsA chemical reduction) with gaseous MM application. The bio-sniffer was calibrated against MM vapor from 8.7 to 11500 ppb with correlation coefficient of 0.977, including a MM threshold (200 ppb) of pathologic halitosis and the human sense of smell level 3.5 (10.0 ppb), with good gas-selectivity based on the MAO-A substrate specificity. As the result of the physiological application, the optical bio-sniffer could successfully monitor the MM level change in breath samples during daytime, which is consistent with the previously reported results.