Article ID Journal Published Year Pages File Type
1172271 Analytica Chimica Acta 2006 9 Pages PDF
Abstract

A new liquid chromatography/atmospheric pressure chemical ionization-tandem mass spectrometry (LC/APCI-MS/MS) method with on-line sample clean-up for the determination of telmisartan in human blood plasma is presented. This technique is compared to a previously introduced enzyme-linked immunosorbent assay (ELISA), where fluorescence is used as detection method. For the LC/MS method applying an internal calibration via a deuterated internal standard, the limit of detection was 0.3 ng/mL, the limit of quantification was 0.9 ng/mL and the linear range extended from 0.9 to 1000 ng/mL. Forty-eight plasma samples from four healthy volunteers were analyzed in a pharmacokinetic study to obtain data for the method comparison. As a result, these two new and independent analytical methods for the determination of telmisartan in human blood plasma proved to yield comparable results for the amount of analyte.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,