Abstract

The corrosion behavior of CoCrMo implants with rough titanium coatings, applied by different suppliers by either sintering or vacuum plasma spraying, has been evaluated and compared with uncoated material. The open-circuit potential, corrosion current and polarization resistance were determined by electrochemical techniques. The Co, Cr and Ti ions released from the samples into the electrolyte during a potentiostatic extraction technique were analyzed using ICP-MS.The Ti coatings from the different suppliers showed a different porous morphology, and the implants exhibited a distinct corrosion activity, underlining the importance of the coating process parameters. Among the titanium coated samples, the one with the sintered overcoat turned out to be the most resistant. Yet, on an absolute scale, they all showed a corrosion resistance inferior to that of uncoated CoCrMo or wrought titanium.

Related Topics
Physical Sciences and Engineering Chemical Engineering Bioengineering