Article ID Journal Published Year Pages File Type
17949 Enzyme and Microbial Technology 2008 5 Pages PDF
Abstract

A microbial biosensing system for the measurement of methane has been developed using an immobilized mixed culture of Pseudomonas aeruginosa and Klebsiella sp. together with a dissolved oxygen (O2) sensor. When methane was applied to the biosensing system, the dissolved O2 content decreased until a steady-state was reached. The biosensing system response depends linearly on methane concentration between 1.0 and 5.0% (v/v) with a detection limit of 0.3% (v/v) (S/N = 3) and a 100-s response time. Phosphate buffer (pH 7.0, 25 mM) and room temperature (20–25 °C) were chosen as the optimum working conditions. Some parameters including pH, temperature, operational and storage stability were studied in detail for characterization of the biosensing system.

Related Topics
Physical Sciences and Engineering Chemical Engineering Bioengineering
Authors
, , , , , ,