Article ID Journal Published Year Pages File Type
180723 Electrochemistry Communications 2010 4 Pages PDF
Abstract

In this paper, a microchip-based sandwich-type aptasensor is developed for the detection of human thrombin. The SH-aptamer/thrombin/alkaline phosphatase-functionalized aptamer (ALP-aptamer) system was constructed in the microfluidic channels. And the substrate solution containing 4-aminophenyl phosphate (p-APP) was introduced to the microchannels for the end-column electrochemical detection. The on-chip aptasensor has a broad linear response range of 1–100 pM with a detection limit of 1 pM, which shows high sensitivity and specificity. The system was then applied to detect thrombin in human serum sample. Therefore, the on-chip aptasensor has a great promise for detecting and screening ultratrace levels of biomarkers in the complex matrices.

Related Topics
Physical Sciences and Engineering Chemical Engineering Chemical Engineering (General)
Authors
, , , , , ,