Abstract

PurposeTo evaluate the liver-to-muscle signal intensity and R2* methods to gain a transferable, clinical application for liver iron measurement.Materials and MethodsSixteen liver phantoms and 33 human subjects were examined using three 1.5-T MRI scanners from two different vendors. Phantom-to-muscle and liver-to-muscle signal intensity ratios were analyzed to determine MRI estimated phantom and hepatic iron concentration (M-PIC and M-HIC, respectively). R2* was calculated for the phantoms and the liver of human subjects. Seven patients' biochemical hepatic iron concentration was obtained.ResultsM-PIC and R2* results of three scanners correlated linearly to phantom iron concentrations (r=0.984 to 0.989 and r=0.972 to 0.981, respectively), and no significant difference between the scanners was found (P=.482 and P=.846, respectively) in vitro. The patients' R2* correlated linearly to M-HIC of the standard scanner (r=0.981). M-HIC values did not differ from those obtained from the biopsy specimens (P=.230). The difference in M-HIC was significant, but the difference in R2* was not significant between the scanners (P<.0001 and P=.505, respectively) in vivo.ConclusionBoth methods, M-HIC and R2*, are reliable iron concentration indicators with linear dependence on iron concentration in vivo and in vitro. The R2* method was found to be comparable among different scanners. Transferability testing is needed for the use of the methods at various scanners.