Article ID Journal Published Year Pages File Type
1807475 Magnetic Resonance Imaging 2007 10 Pages PDF
Abstract

Coronary magnetic resonance angiography (MRA) acquired using steady-state free precession (SSFP) sequences tends to suffer from image artifacts caused by local magnetic field inhomogeneities. Flow- and gradient-switching-induced eddy currents are important sources of such phase errors, especially under off-resonant conditions. In this study, we propose to reduce these image artifacts by using a linear centric-encoding (LCE) scheme in the phase-encoding (PE) direction. Abrupt change in gradients, including magnitude and polarity between consecutive radiofrequency cycles, is minimized using the LCE scheme. Results from numeric simulations and phantom studies demonstrated that signal oscillation can be markedly reduced using LCE as compared to conventional alternating centric-encoding (ACE) scheme. The image quality of coronary arteries was improved at both 1.5 and 3.0 T using LCE compared to those acquired using ACE PE scheme (1.5 T: ACE/LCE=2.2±0.8/3.0±0.6, P=.02; 3.0 T: ACE/LCE=2.1±1.1/3.0±0.8, P=.01). In conclusion, flow- and eddy-currents-induced imaging artifacts in coronary MRA using SSFP sequence can be markedly reduced with LCE acquisition of PE lines.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Condensed Matter Physics
Authors
, , , , , ,