Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1807973 | Magnetic Resonance Imaging | 2006 | 13 Pages |
Abstract
Although diffusion tensor imaging (DTI) shows great potential for the diagnosis of a variety of pathologies, no consensus for an appropriate assessment standard of DTI exists. This study examined the feasibility of using water-filled arrays of glass capillaries to construct a DTI phantom suitable for making repeated and reproducible measurements required in a quality assessment program. Three phantoms were constructed using arrays of capillaries with three inner diameters (23, 48, and 82 μm). Data were acquired using DTI protocols; the fractional anisotropy (FA), mean apparent diffusion coefficient (ãADCã) and principal eigenvectors of the diffusion tensors were calculated. This study demonstrated four results: (1) echo-planar images show that susceptibility within the capillary arrays does not lead to substantial differences in precessional frequency in regions containing the arrays and neither do the regions show noticeable image distortion; (2) principal eigenvectors of the diffusion tensors agree to within <10.3° of the array orientations; (3) mean FA values (0.18-0.50) and ãADCã values (1.40-1.93Ã10â3 mm2/s) within specified regions of interest are in general agreement with simulations after a simple noise correction; and (4) these array performance characteristics are observable using a typical clinical DTI protocol.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Condensed Matter Physics
Authors
Nathan Yanasak, Jerry Allison,