Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
181112 | Electrochemistry Communications | 2009 | 4 Pages |
Abstract
We report on a novel glucose biosensor based on the immobilization of glucose oxidase (GOx) on a Prussian blue modified nanoporous gold surface. The amperometric glucose biosensor fabricated in this study exhibits a fast response and the very low detection limit of 2.5 μM glucose. The sensitivity of the biosensor was found to be very high, 177 μA/mM; the apparent Michaelis–Menten constant is calculated to be 2.1 mM. In addition, the biosensor has good reproducibility and remains stable over 60 days. The anti-interference ability of the biosensor was also assessed, showing little interference from possible interferents such as ascorbic acid (AA), acetaminophen (AP) and uric acid (UA).
Related Topics
Physical Sciences and Engineering
Chemical Engineering
Chemical Engineering (General)
Authors
Asieh Ahmadalinezhad, A.K.M. Kafi, Aicheng Chen,