Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1826582 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2010 | 7 Pages |
Abstract
In this paper, we introduce a direct analytical mathematical method for calculating the geometrical and absolute total efficiencies of gamma NaI scintillation detectors using an arbitrarily positioned point source. Generalizing, the geometrical and absolute total efficiencies are extended to include coaxial and off-axis sources with plane and volumetric shapes. The efficiencies are deduced into straightforward mathematical expressions. The current theoretical and the published efficiency values are in good agreement.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
A. Hamzawy,