Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1883390 | Radiation Physics and Chemistry | 2014 | 8 Pages |
Abstract
The objective of this study was to evaluate the entrance surface air kerma in pediatric chest radiography. An evaluation of 301 radiographical examinations in anterior-posterior (AP) and posterior-anterior (PA) (166 examinations) and lateral (LAT) (135 examinations) projections was performed. The analyses were performed on patients grouped by age; the groups included ages 0-1 y, 1-5 y, 5-10 y, and 10-15 y. The entrance surface air kerma was determined with DoseCal software (Radiological Protection Center of Saint George׳s Hospital, London) and thermoluminescent dosimeters. Two different exposure techniques were compared. The doses received by patients who had undergone LAT examinations were 40% higher, on average, those in AP/PA examinations because of the difference in tube voltage. A large high-dose “tail” was observed for children up to 5 y old. An increase in tube potential and corresponding decrease in current lead to a significant dose reduction. The difference between the average dose values for different age ranges was not practically observed, implying that the exposure techniques are still not optimal. Exposure doses received using the higher tube voltage and lower current-time product correspond to the international diagnostic reference levels.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
L. Porto, N. Lunelli, S. Paschuk, A. Oliveira, J.L. Ferreira, H. Schelin, C. Miguel, V. Denyak, C. Kmiecik, J. Tilly, H. Khoury,